Buffers for LC

What Buffers to use?

Introduction

In reversed-phase HPLC, the pH of the eluent can significantly influence the separation of components. Buffers are required when the sample contains ionic or ionisable analytes. Without a buffer, poor peak shape and variable retention may result.

In general, for acids, increasing eluent pH leads to increased ionisation and a decrease in retention (see Figure 1). For bases, decreasing pH results in greater ionisation and decreased retention. For robust methods, separations should be developed at a pH where retention is least affected by pH change (see robust pH zones in Figure 1). 

Buffer pH-K wert
Fig 1

 

Choice of Buffer pH

Optimum buffering capacity occurs at a pH equal to the pKa of the buffer. In general, the effective pH range for a buffer is within 1 pH unit of it’s pKa. Table 1 indicates the pH range of some common buffers. The buffer pH should be selected so that it is at least 1 pH units from the pKa of the analyte. This will ensure that the analytes are 100% ionised or 100% non-ionized, for reproducible retention. 

Further Considerations

At pH >7, phosphate buffers accelerate the dissolution of silica. For optimum column lifetime, organic buffers (eg pyrrolidine) should be used for pH above 8. Buffer concentrations in the 10 - 25 mM range are recommended (maximum 50mM). Fresh buffer solutions should be prepared regularly and pre-column filters should be used to protect columns. Buffers should be flushed from the column prior to storage.

Buffers for LC-MS

For LC-MS analyses, volatile buffers or additives are preferred, in order to minimise MS ion suppression and maintain sensitivity. In addition, non-volatile buffers such as phosphate may lead to contamination of the ion source. Buffer concentrations should be as low as possible eg 10 to 20mM. Ammonium salts are more volatile than those of Na+ or K+. TFA should be avoided with electrospray LC-MS as it reduces sensitivity. Formic acid or acetic acid (0.01 to 1% v/v) are preferred. For higher pH applications, ammonium hydroxide is recommended.

Buffer or Additive

pKa

pH Range

UV Cut-off(nm)

MS Compatible

Comments

TFA

0.3

-

 

yes

can suppress ionisation Usage 0.02 -0.1%

Phosphate 1

2.1

1.1 - 3.1

 

no

 

Phosphate 2

7.2

6.2 - 8.2

 

no

 

Phosphate 3

12.3

11.3 - 13.3

 

no

 

Formic acid

3.8

-

 

yes

 

Ammonium Formate

3.8

2.8 - 4.8

 

yes

 

Acetic acid

4.8

-

 

yes

 

 Ammonium Acetate

4.8

3.8 - 5.8

 

yes

 

Citrate

3.1, 

2.1 - 4.1

 

 

 

Citrate

 4.7

3.7 - 5.7

 

 

 

Citrate

5.4

4.4 - 6.4

 

 

 

Ammonium Bicarbonate

10.3 (HCO3)
9.2 NH4+

8.2 - 11.3

 

yes

 

Bicarbonate

10.3

9.3 - 11.3

 

 

 

Ammonia / ammonium hydroxide

9.2

8.2 - 10.2

 

Yes

 

Borate

9.2

8.2 - 10.2

 

no

 

CAPSO

9.7

8.7 - 10.7

 

no

 

Glycine

2.4 , 9.8

8.8 - 10.8

 

no

 

 Tris

8.3

7.3 - 9.3

 

 

 

1-Methylpiperidine

10.1

9.1 - 11.1

 

yes

 

CAPS

10.4

9.5 - 11.5

 

no

 

Diethylamine

10.5

9.5 - 11.5

 

 

 

 Pyrrolidine

11.1

10.1 - 12.1

 

yes

 

Triethylamin acetate salts (TEA)

10.7

9.7 -11.7

 

yes

 

pH of Acetate Buffer at 20 C

20 mM CH3COOH

10

9

8

7

6

5

4

3

2

1

0

20 mM CH3 COONa

0

1

2

3

4

5

6

7

8

9

10

pH

3.2

3.8

4.1

4.3

4.6

4.7

4.8

5.0

5.3

5.6

6.8

pH of Formate Buffer at 20 C

50 mM HCOOH

10

8

6

4

2

1

0.5

0.2

0.1

0

50 mM HCOONH4

0

2

4

6

8

9

9.5

9.8

9.9

10

pH

2.5

3.1

3.5

3.8

4.2

4.7

4.9

5.4

6.1

6.4

 

I

© by LCC 1995 -2016 All rights reserved
 Chromatographyshop is a division of LCC Engineering & Trading GmbH, CH-4622 Egerkingen, Switzerland.
Collaboration partners in various countries
Our Company Profile and Summary Product List 2015
Unser Firmenprofil und Produktzusammenfassung 2015
      Our Terms and Conditions

Sample Preparation Tools

[Sample  Preparation]
[MatrixEx]
[QuEChERS]
[SPE]
[Syringe Filters]
[Inert RSA  Glass Vials]
[Flash Chromatography]

General Issues

[Chromatogr. Shop Home]
[Newsletter Archive]
[Our Services & Company]
[Terms & Conditions]
[Mobile Phone]

 Bonuses for you

[Special Deals for you]

Seminars & Workshops

[Knowledge Education]
[Seminars & Workshops]

Analytical Scale Chromatography

[Analytical Columns]
[ACE & ACE  Excel Columns]
[ACE Ultra Core]
[Microsolv Cogent Column]
[Capcell Shiseido]
[Daicel / Chiral Technologies]
[GL Science  Inertsil]
[Hichrom Columns]
[Imtakt Column]
[Kromasil Akzo Nobel]
[Mitsubishi MCI-Gel]
[Merck Lichro...]
[Macherey &  Nagel]
[Princeton  Chromatography]
[PolyLC  Columns]
[SieLC Primesep]
[Capcell Shiseido]
[Ultron by Shinwa]
[Welch  Materials]
[Zirchrom]
[Agilent Zorbax]

Knowledge Tools for Chromatographers

[Knowledge Education]
[Definition of  Terms Used]
[Separation Mechanism]
[Solvent  Selection]
[Buffers for  LC.]
[Column &  Material Care]
[Prep & Process Columns]
[Analytical  Columns]
[HPLC- to-UHPLC]

Resin based Systems (Analytical to Process scale)

[Bioseparation Biopurification]

Special Tools to improve data quality, reproducibility and productivity

[Inert RSA  Glass Vials]

Method & Material  Choices  1

[Knowledge Education]
[Sample  Preparation]
[Reverse_Phase_LC]
[Adsorption-LC RPC & SPE]
[AQ Columns]
[Phenyl Bond Columns]
[FPF_Columns]
[HILIC & Aq. Norm. Phase]
[Electrostat. Repul. HILIC]
[Normal Phase Chromatography]

Process Scale   Chromatography Equipment

[Process  Chromatogr.]
[Process Chromat.Systems]
[Flash &  MPLC Pumps]
[HPLC  Pumps]
[Special  Pumps]
[Fraction Collectors]
[Stainless Steel Columns]
[Glass  Columns]
[Glass  Cartridges]
[Plastic Cartridges]
[Manual  Valves]
[Motor  Valves]
[Efficiency  Tools]

Method & Material Choices 

[Knowledge Education]
[Multi  Mode LC]
[Chiral  Chromatography]
[Ion Exchange  & Ion LC]
[Hydrophobic  Interaction LC.]
[Affinity  LC]
[Size Exclusion Chromatogr.]
[SuperCritical Fluid LC]
[Proteomics]
[Metabolomics]
[Process  Scale LC]
[Kamasutra Inspired Tech]